| Item No | 4028 Series |
| Item Name | 2% Accurate, Precision Power Sensors |
| Frequency Range | See Chart |
| RF Power | See Chart |
| VSWR | 1.05:1 max. |
| Insertion Loss | 0.05 dB max. (with female 7-16 DIN connectors) |
| Directivity | 28 dB (min.) |
| Connectors | A Series: Customer specified from QC connector, appropriate for frequency and power B Series: 1 5/8” EIA Flanged C Series: 3 1/8” EIA Flanged |
| Impedance Ω | 50 (nominal) |
| Max Allowable Terminating VSWR | 2.00:1 |
| Calibration Power Level | 3.5 kW, typical |
| Calibration Technique | Frequency-specified calibration factors stored in nonvolatile memory in each sensor. |
| Calibration Cycle | 1 year (nominal) |
| Accuracy | ± 2.0% (2 ó) |
| Accuracy - Outside Cal Power and Frequency | ± 4.0% (2 ó) |
| Accuracy - RFL | = Fwd accuracy + (Fwd Power/ 10^{directivity/10}) |
| Accuracy - VSWR | Calculated from FWD and RFL power (VSWR = [1+Sqrt(Pr/PF)] / [1-Sqrt(Pr/Pf) / [1-Sqrt (Pr/Pf)]) |
| Sampling Rate | 2 readings / second (nominal) |
| Operating Power | Supplied by power meter via sensor cable |
| Operating Temperature | 0 to 50 °C (32 to 122 °F) (derate accuracy outside 25 ± 10 °C) |
| Storage Temperature | -20 to +70 °C (-4 to + 158 °F) |
| Humidity | 95% non-condensing |
| Altitude | 10,000 ft. (3,000 m) max. |
| Repeatability - Multiple Measurements-Single Sensor | 0.3% (95% c.l.) (with female 7-16 DIN connectors) |
| Dimensions | (L x W x H) A Series: 4.7” x 3.2” x 3.8” (120 x 82 x 97 mm) -without connectors B Series: 6.8” x 3.5” x 4.8” (171 x 89 x 121 mm) C Series: 8.0” x 5.2” x 6.4” (203 x 131 x 162 mm) |
| Weight | A Series: 1.7 lbs (.8 kg) B Series: 3.3 lbs (1.5 kg) C Series: 7.3 lbs (3.3 kg) |